SPM300

SPM300

SPM300 은 Muller Matrix 방식의 High speed IL PDL meter 입니다. 특히 AWG , CWDM filter 측정에 최적화 되어 있습니다. Santec TLS 와 Agilent TLS와 Trigger interface를 사용해서 측정 속도를 극대화 시킨 제품입니다.

Key Features

AWG,VOA etc module measurement(IL,PDL) system
Including a AWG analysis software.(TE,TM Analysis)
High speed measurement (above 5nm/Sec, 5pm step).
Large logging memory(40000 point per channel)
Mueller matrix method(4state , 6 state)
One chip measurement time is below 3 minutes.
Real time monitoring and compensation of source power variation
Low cost solution
Scalable structure graphic user interface
GPIB,TCP/IP, USB, RS232C Remote Interface

Data Sheets

SPM300

Parameter  
Wavelength range1270~1630nm(depend on TLS)
Measurement resolution1pm(depend on TLS)
IL measurement accuracy±0.05dB( 0~20dB device IL)
±0.10dB( 20~40dB device IL)
±0.20dB( 40~60dB device IL)
IL measurement range10dBm ~ -70dBm
IL measurement repeatability±0.01dB
PDL measurement accuracy±0.05dB( 0~20dB device IL)
±0.10dB( 20~40dB device IL)
PDL measurement repeatability±0.015dB
PDL measurement range10dBm ~ -60dBm
Minimum Acquisition time for single wavelength1 msec
Minimum measurement time4 Scanning time(Muller matrix)
* ( 1 + 0.02*# of Channels)
PSG Insertion Loss1.0dB typical
PSG Wavelength dependent Loss0.3dB
PSG State dependent loss(64-States)< 0.1dB
Maximum number of channels256
Analysis Parameters Define 
Center WavelengthDefined as the center of the 3dB passband
Wavelength OffsetDefined as the difference between the ITU Wavelength and Center wavelength
Polarization Dependent WavelengthDefined as the difference between the center wavelengths for TE and TM Polarization
Mininum Insertion LossDefined as the minimum loss within the ITU passband for TE and TM Polarization(all polarization states)
Maximum Insertion LossDefined as the maximum loss within the ITU passband for TE and TM Polarization(all polarization states)
RippleDefined as the difference between Minimum Insertion Loss and Maximum Insertion Loss within ITU passband for all polarization states
Maximum Polarization Dependent LossDefined as the maximum difference in IL between TE and TM polarization within the ITU passband.
Center Polarization Dependent LossDefined as the maximum difference in IL between TE and TM at the center wavelength
UniformityDefined as the maximum difference between maximum Insertion Loss across all channels
Passband 1dB(3dB,10dB,20dB)Defined as the minimum effective bandwidth 1dB (3dB , 10dB , 20dB) down from peak transmission
Adjacent Channel CrosstalkDefined as the difference between maximum loss in passband and maximum transmission in adjacent channel passbands
Non Adjacent CrosstalkDefined as the difference between maximum loss in passband and worst case maximum transmission in all non-adjacent channel paassbands.
Total CrosstalkDefined as the cumulative sum of all adjacent channel and non-adjacent channel crosstalk values , at worst case polarization.
Key FutureSpecification
External ControlGPIB(Optional)IEEE 488.2 (Compatible)
TCP/IPNormal TCP/IP
USBUSBTMC Spec
RS232C9 pin D-sub , 115200 baudrate
Dimension5U , Fullrack size
AC Power Input90~240V , 50 /60 Hz
Operating Temp.10 to 40 ℃
Storage Temp0 to 60 ℃